User's Manual

19
Ensure that there is no other Trumpet connected to the power
plug and to the USB plug in the surroundings, otherwise the
probes could pair to the wrong device.
In case of any problemi it is possible to turn off the probes manually by
keeping the button pressed for more than 10 seconds. They are however
turned off every time that Trumpet is not in use by Maestro, unless they are
in charge position.
The following table summarize the possible probes LED statuses:
Color
Meaning
Turned off
Probes turned off
Blue blinking
Probes searching for the Trumpet
Orange fixed
Probes connected to the Trumpet
Green (short)
Probes charge start
2.2.3 Equalization of active speaker (REM)
During the speaker equalization procedure, the headset with the REM
probes must be placed in front of the speaker, at the same distance and in
the same position as will be occupied by the patient's head when the test is
actually conducted. For best results, it is advisable to fit the headset directly
to the patient before running the equalization procedure. The distance of 1
m from the speaker is the optimum distance both for equalization purposes
and for conducting the REM test. Make certain that there are no
obstructions (e.g. hair) covering the microphones.
For more details, refer to the manual entitled Maestro Fitting and hearing
aid test functionalities User Manual.
2.2.4 Probe tube calibration (REM)
The position for calibration of the probe microphones is indicated below
(refer to Maestro Fitting and hearing aid test functionalities User
Manual for further information on how to proceed with calibration). The tip
of the probe tube must be inserted into the relative guides, as near as
possible to the reference microphone, and the headset placed at around 1
metre from and in front of the active speaker.