User's Manual

28
The response in the auditory canal can also be measured using the earmold
of the patient's hearing aid, rather than a disposable tip. Likewise in this
instance, the probe tube must first be inserted in the auditory canal of the
patient as described in Heading 2.3.1, following which the mold, connected
previously to the insert transducer, can be placed in the ear. Here again,
special care must be taken during the operation to avoid displacing the
probe tube inserted previously from its correct position. An example of the
final set-up is illustrated in the following image:
Electing to use the patient's earmold for acquisition of the measurement in
the ear, there are two distinct configurations possible for measuring the
response in the coupler.
The first configuration, identified as HA1, requires the use of the adapter
(1). As with the disposable tip, connect the adapter (1) to the coupler, and
the earmold to the insert transducer. Finally, attach the earmold to the
adapter, using adhesive insulating putty to ensure that the two parts are
securely connected with no gaps left between them. The correct position is
shown in the following image: