Use Instructions

16 Safety-related checks
The Invia Ease pump is a device in protection class II (EN IEC 60601-1), the safety-related
checks are confined to visual inspection of the housing and charger for damage.
This check must be performed prior to each use.
If any damage or safety defects are observed, contact the Medela Customer Service
or the system provider.
Devices of protection class II do not have a protective earth conductor therefore there
is no need to check the earth leakage current.
The Invia Ease pump enclosure is made entirely of insulated material. Tests of the enclosure
leakage current using common measuring instruments will therefore not reveal measurable
values.
Even when suctioning a conductive fluid until the overflow protection device activates,
measurements of the patient leakage current using common measuring instruments will
not reveal measurable values.
The Invia Ease pump does not have patient circuits or functional earth connections.
The Invia Ease pump and charger should be maintained and repaired throughout
its service life in compliance with the Medela service procedures.
Service work may only be carried out by Medela’s authorized personnel.
To request maintenance service contact the Medela Customer Service.
17 Maintenance and service
Invia Ease pump, charger, handle, IV pole/bed holder and carrying case cannot
be sterilized.
The following products are sterile and single use only:
Invia Ease canister and tubing
These are single use products that should be disposed of after use.
Do not reuse.
If reused, the performance of the product may deteriorate, cross-contamination may occur.
18 Sterility requirements
Sterile products: Do not use the sterile products if the sterile packaging is damaged,
was opened prior to use or has expired.
WARNING
EN
Safety-related checks/Maintenance and service/Sterility requirements | 49