Use Instructions

To disinfect the Invia Ease pump and the accessories (charger, handle and IV pole
fixation):
1. Wear suitable protection (clothing, gloves, face mask and goggles) according to the facility
guidelines.
2. Apply the disinfectant agent in accordance with instructions from the manufacturer.
Pay particular attention to the edges, narrow corners, and bottom side.
3. Leave the disinfection agent on for the time recommended by the manufacturer.
4. Thoroughly clean the surface, all edges, housing niches, corners, and bottom side.
5. Wipe dry or air dry according to the instructions from the manufacturer.
6. If needed, repeat step 2–5.
7. Dispose contaminated material in accordance with local environmental guidelines.
For detailed cleaning instructions contact the Medela Customer Service.
19.3 Drying
After the manual disinfection, store the product on a clean dry cloth and let it dry.
Alternatively, follow the instructions provided by the disinfectant manufacturer.
19.4 Inspection
After reprocessing according to the steps described above, the device and its components
must be inspected for signs of degradation that may limit the useful life and/or performance
of the device, such as the following: visible corrosion, mechanical wear, abrasion, damage,
or deformation. Furthermore, check if the silicone O-ring is still in place (if applicable).
Discard the device if any signs of degradation are evident.
19.5 Storage
Store the product dry and dust-free.
EN
Cleaning, disinfection and storage/Disposal | 51
The Invia Ease pump is made from various metal and plastic parts, which should be
disposed of in accordance with the local environmental guidelines.
Before disposal of the pump, the rechargeable battery and electronics must be removed.
For instructions, contact the Medela Customer Service.
20 Disposal
NOTICE
Invia Ease products should be handled and disposed of in accordance with the local
environmental guidelines.
The Invia Ease pump or the accessories must not be disposed of with the household refuse.