Use Instructions

EN
International regulations | 53
Guidance and manufacturer’s declaration – electromagnetic emissions
This Invia Ease pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Invia Ease pump should assure that it is used in such an
environment.
Emission tests Compliance Electromagnetic environment – guidance
RF Emissions
CI SPR 11
Group 1
The Invia Ease pump uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
Class B
The Invia Ease pump is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Not
applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment
– guidance
Electrostatic
discharge ( ESD )
IEC 61000-4-2
± 8 kV
contact discharge
± 15 kV
air discharge
± 8 kV
contact discharge
± 15 kV
air discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material,
the relative humidity
should be at least 5 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power supply
lines
± 2 kV
for power supply
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Line-to-line
± 0,5 kV, ± 1 kV
Line-to-line
Mains power quality
should be that of a typical
commercial or hospital
environment.
Guidance and manufacturer’s declaration – electromagnetic immunity