User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
![](/manual/invivo/3gecg/user-manual-english/images/img-118.png)
4‐8AlarmsExpressionMR400InstructionsforUse
Managing Alarm Functions
WARNINGS
• Always respond promptly to any alarm condition; otherwise, treatment of the patient
could be delayed.
• You should ensure that the current alarm preset is appropriate prior to use on each
patient. Failure to do so may cause a lapse in patient monitoring.
• Setting the alarm limits to extreme values can render alarm monitoring useless. Also,
potential hazard can exist if different alarm monitoring settings are used for the same
or similar equipment in any single patient care unit. Ignoring these restrictions may
cause a lapse in patient monitoring.
TheMR400canbesettoprovidevisualalarmsignalsonly,orbothvisualandaudiblealarm
signals.Theaudiblealarmstateswi llhavenoeffectonanyofthevisualalarmsdisplayedbythe
MR400(orIP5,ifequipped).
AllsettingsintheAlarmsmenu(exceptAlarmSound)
canbestoredandrecalled.
Restrictionstoalarmpresetsarenotprovidedduetothefastpacedworkflow,shortaverage
durationtimeofMRIscans,differentpatienttypesscheduledforMRIprocedures,anddirect
patientsupervisionprovidedduringMRIprocedures.
Note
If the MR400 is networked to an IP5, alarm indications occur at both the MR400 and the IP5; and
control of alarms, including sound level, is localized to the device (MR400 or IP5) indicating the
alarm condition.
Showing or Hiding Current Alarm Limits
CurrenthighandlowalarmlimitsettingsaredisplayedintheVSboxofeachmonitored
parameterbydefault(exceptbellows‐derived respiration).
Tocontrolthedisplayfunctionforthealarmlimitsettings
Step Action
1PresstheSetupkeyandthentheAlarmskey.
TheAlarmsmenuappears.Currentsettingsaredisplayed.