User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
4‐28AlarmsExpressionMR400InstructionsforUse
Missing heart rate data ECG, SPO2, P1
(or P2)
The heart rate data, once present,
can no longer be produced.
Flashing yellow dashes (- - -) in the heart rate
numeric in the ECG VS box and in the SPO2
VS box, yellow alarm light, medium priority
alarm tone
Over maximum heart
rate value
ECG, SPO2, P1
(or P2)
The heart rate measurement has
violated the upper parameter
range.
Flashing red heart rate numeric alternating
between OVR and the heart rate value in the
ECG VS box and in the SPO2 VS box, red alarm
light, high priority alarm tone
Under minimum heart
rate value
ECG, SPO2, P1
(or P2)
The heart rate measurement has
violated the lower parameter
range.
Flashing red heart rate numeric alternating
between UND and the heart rate value in the
ECG VS box and in the SPO2 VS box, red alarm
light, high priority alarm tone
Desat SPO2 SPO2 has detected a desaturation
event.
Red flag in ECG alarm flag area, flashing red
SpO2 numeric, red alarm light, high priority
alarm tone
Violated arterial oxygen
saturation value
SPO2 The arterial oxygen saturation
measurement has violated an
alarm limit setting.
Flashing yellow SpO2 numeric in the SPO2 VS
box, yellow alarm light, medium priority alarm
tone
Missing arterial oxygen
saturation data
SPO2 The arterial oxygen saturation
data, once present, can no longer
be produced.
Flashing yellow dashes (- - -) in the SpO2
numeric in the SPO2 VS box, yellow alarm
light, medium priority alarm tone
Apnea CO2, AGENT The apnea time delay setting has
been exceeded.
Red flag in the CO2 or RESP alarm flag area,
Flashing red respiration rate numeric, red
alarm light, high priority alarm tone
Note
Apnea alarm flag appears in either the CO2 or
the RESP alarm flag area, based on the RESP
source.
Violated EtCO2 value CO2, AGENT The end-tidal CO2 measurement
has violated an alarm limit setting.
Flashing yellow EtCO2 numeric in the CO2 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified
Missing EtCO2 data CO2, AGENT The end-tidal CO2 data, once
present, can no longer be
produced.
Flashing yellow dashes (- - -) in the EtCO2
numeric in the CO2 VS box, yellow alarm light,
medium priority alarm tone
Violated FiCO2 value CO2, AGENT The fractional inspired CO2
measurement has violated an
alarm limit setting.
Flashing yellow FiCO2 numeric in the CO2 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified
Missing FiCO2 data CO2, AGENT The fractional inspired CO2 data,
once present, can no longer be
produced.
Flashing yellow dashes (- - -) in the FiCO2
numeric in the CO2 VS box, yellow alarm light,
medium priority alarm tone
Patient Alarm From Condition Indication