User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
4‐34AlarmsExpressionMR400InstructionsforUse
CO2 Low Flow
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Message may appear when the sampling line is initially connected; allow the a few seconds for the
flow to be established. Otherwise, the detected flow rate is 10 percent less than nominal; in this
case, check the sampling line for pinches or obstructions then clear any pinch or replace if
necessary.
If the message persists, contact technical support or authorized service personnel.
Check for CO2 Occlusion
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Detected flow through the sampling line is less than 40 ml/min or the water trap may be full of
fluid:
• Check the sampling line for obstructions and replace it if necessary.
• Check the fluid level in the water trap and replace it if necessary.
If the message persists, contact technical support or authorized service personnel.
CO2 Out Of Range
CO2 alarm flag area, blue
alarm light, INOP alarm tone
The calculation value is greater than the upper CO2 limit. Perform readings to confirm patient’s
physiological condition.
If the message persists, contact technical support or authorized service personnel.
CO2 Sensor Faulty
CO2 alarm flag area, blue
alarm light, INOP alarm tone
CO2 bench detected a hardware or sensor error. Cycle AC mains power.
If the message persists, contact technical support or authorized service personnel.
CO2 Sensor Over Temp
CO2 alarm flag area, blue
alarm light, INOP alarm tone
The CO2 sensor is above the specified operating temperature. Confirm that the MR400 is
operating within the required environmental conditions (see Appendix A); if outside the specified
range, move the MR400 to an area that is within limits.
If the problem persists, stop all monitoring activities and contact technical support or authorized
service personnel.
CO2 Warming Up
CO2 alarm flag area, blue
alarm light, INOP alarm tone
CO2 is warming to operating temperature. Allow the process to complete, about 2 minutes.
CO2 Zero Required
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Zero calibration of the CO2 module is needed; see page 7-15 for details.
Magnetic Field Too High
CO2 alarm flag area, blue
alarm light, INOP alarm tone
The gauss limit has been exceeded, and the AGENT option can no longer function. Position the
MR400 per the product labeling; see page 3-2.
HW Fail
CO2 alarm flag area, blue
alarm light, INOP alarm tone
CO2 module was not found during initialization. Cycle AC mains power.
If the message persists, contact technical support or authorized service personnel.
Low O2
AGENT alarm flag area, blue
alarm light, INOP alarm tone
The inspired O2 value has dropped below 18.0
Motor Speed Error
CO2 alarm flag area, blue
alarm light, INOP alarm
tone
The MR400 is too close to the MR magnet.
Ensure that the MR400 has been positioned correctly; see page 3-2. If the message persists, contact
technical support or authorized service personnel.
Message, Location,
Indication
What to Do