ExpressionMR400InstructionsforUseAlarms437
NIBP
Message, Location,
Indication
What to Do
Communication Error
NIBP alarm flag area, blue
alarm light, INOP alarm tone
An internal NIBP error has occurred. Discontinue use of NIBP and contact technical support or
authorized service personnel.
Deflation Timeout
NIBP alarm flag area, red
alarm light, high priority
alarm tone
Cuff deflation has timed out; displayed if the NIBP cuff deflation period is greater than 80 seconds
(neonatal patient type) or is greater than 150 seconds (adult and pediatric patient types):
Check the patient.
Check for proper cuff size and placement.
Check the cuff and hoses for pinching.
If the message persists, contact technical support or authorized service personnel.
Hardware Error
NIBP alarm flag area, blue
alarm light, INOP alarm tone
NIBP hardware failure detected. Discontinue use of NIBP and contact technical support or
authorized service personnel.
Inflation Timeout
NIBP alarm flag area, blue
alarm light, INOP alarm tone
Cuff inflation has timed out:
Check the patient.
Check the cuff and hoses for pinching or leaks.
If the message persists, contact technical support or authorized service personnel.
Measurement Failed
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The NIBP measurement has failed:
Check the patient.
Check for proper cuff size and placement.
Check the cuff and hoses for pinching or leaks.
If the message persists, contact technical support or authorized service personnel.
Measurement Timeout
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The NIBP measurement has timed out:
Check the patient.
Check for proper cuff size and placement.
Check the cuff and hoses for pinching or leaks.
Check the cuff condition and placement on the patient.
Module Not Calibrated
NIBP alarm flag area, blue
alarm light, INOP alarm tone
NIBP is not calibrated. Contact technical support or authorized service personnel.
Over Pressure
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The allowed pressure for the type of patient has been exceeded:
Ensure that the patient is immobilized and not applying pressure to the cuff.
Check the cuff condition and placement on the patient.
Make sure that the hose is not pinched.
If the message persists, contact technical support or authorized service personnel.
Pressure Correction
NIBP alarm flag area, blue
alarm light, INOP alarm tone
A pressure correction error has been detected:
Ensure that the patient is immobilized and not applying pressure to the cuff.
Check the cuff condition and placement on the patient.
Residual Pressure
NIBP alarm flag area, blue
alarm light, INOP alarm tone
Residual pressure remains in the NIBP system. Disconnect the NIBP hose from the patient
connection panel and then restart the procedure.