User Manual

Table Of Contents
ExpressionMR400InstructionsforUseMonitoringECG513
CAUTION
ECG lead cable clips should not be placed on the patient’s extremities.
ToattachtheECGleadcabletothewECGmoduleandthentotheQuadtrodeelectrode
contacts
AAMI ECG Lead Cable
Connections
IEC ECG Lead Cable
Connections
AAMI ECG Lead Cable Clip
Designator / Color
IEC ECG Lead Cable Clip
Designator and Color
Associated Limb
RA / White R / Red Right arm
RL / Green N / Black Right leg
LA / Black L / Yellow Left arm
LL / Red F / Green Left leg
Step Action
1InserttheconnectoroftheECGlead
cableintothecableportonthewECG
module.
CAUTION
When inserting or removing the lead
cable, only use the connector as a
finger-hold; never pull or apply
excessive force to the wires.