514MonitoringECGExpressionMR400InstructionsforUse
2 DependingupontheECGleadcabletype,attachtheclipstothe
Quatrodeelectrodecontacts,asshownintheappropriate
connectiondiagramonpage512.Squeezeeachclipopenthen
placetheclipontotheelectrodecontactandrelease.
CAUTION
When inserting or removing the clip leads, use the clip as the finger-
hold; never pull or apply excessive force to the wires.
3IfusingastandardQuadtrodeelectrode,securetheleadcable
wiresusingtheleadretainer.
Seepage53forthelocation.
4CheckthebatteryindicatorsonthewECGmoduletoensurethat
enoughchargeexistsinatleastoneoftheinstalledbatteries:
Greenbatteryindicator=Charge sufficient;proceed
tostep6.
•Redbatteryindicator=Chargelow;proceedtostep5.
Seepage29fordetails.(Also,youcanreferencethe status
informationpane;seepage216.)
5 Accordingtotheredbatteryindicator(s)presentonthewECG
module,insertachargedmodulebatteryintothecorresponding
batterybay(s)
andthenrecheckthebatteryindicator(s)toensurea
sufficientchargebeforeproceeding;seepage124.
6CheckthenetworkchannelindicatoronthewECGmoduleto
ensurecommunicationisestablishedwiththeMR400:
Steady=Goodcommunication;proceedtostep8.
Flashing=Nocommunication;proceedtostep7.
Seepage
210fordetails.(Also,youcanreferencethestatus
informationpane;seepage216.)AninoperativeECGparameteror
wECGmoduleisindicatedbyabsenceofanECGwaveformanda
simultaneousLeadFailalarm.
7EnsurethatthewECGmoduleiswithin9.1m(30feet)ofthe
MR400,
inthesameMRIroomorinthesameshieldedroom,andis
settothesamewirelessnetworkchannelusedbytheMR400;see
page129.