ExpressionMR400InstructionsforUseMonitoringECG525
Tochangetheheartratealarmlimitsettings
Note
See chapter 4 for detailed alarm limit setting instructions and options.
7Enter button
8 Current adjustment
9 Extreme Bradycardia delta value
10 Extreme Bradycardia alarm
setting
11 Alarm limit, minimum
12 Lower alarm limit setting
13 Upper alarm limit setting
14 Alarm limit, maximum
15 Extreme Tachycardia alarm setting
16 Extreme Tachycardia delta setting
Step Action
1 SelectthealarmlimitsettingsintheECGVSbox.
TheHRAlarmLimitsmenuappears.Currentsettingsaredisplayed.
2 SelecttheLow,High,∆ExtrBrady,or∆ExtrTachybutton.
Theselectedbuttonwillbehighlightedandthecurrentadjustment
willbedisplayed.
3Usingthekeypad,
ortheincrement,decrement,orOffbuttons,
enterthedesiredsetting.
Thecurrentadjustmentwillreflectthesetting.
4PresstheEnterbuttontosavethesetting.
Thealarmlimitsettingisupdated.
5Tochangeotheralarmlimitsettings,repeatsteps2,3,and4.
Thecurrentadjustmentwillreflectthechange.