ExpressionMR400InstructionsforUseMonitoringECG533
Note
If Patient Type is set to Neo then Pediatric ECG is set to On and locked. When Patient Type is
changed to Adult, Pediatric ECG will be set to off and unlocked.
TocontrolpediatricECGfiltering
T-Wave Suppression
AllowsyoutoreducetheTwaveamplitudewhenextremelylargeduetothe
magnetohydrodynamiceffect(MHD),whichcanpreventgating.Useforaccurategatingwhenan
unusuallyhighTwaveamplitude,relativetotheRwaveamplitude,isproduced.
Thefollowingoptionsareavailable:
•Off(Default)
•On
Note
T-Wave Suppression is unavailable when Filter Mode is set to Monitor; see page 5-31.
Pediatric ECG Patient Type Condition
Off Adult Unlocked
On Pediatric Unlocked
On Neonatal Locked
Step Action
1PresstheSetupkeyandthentheMonitorkey.
TheMonitorSetupmenuappears.Currentsettingsaredisplayed.
2LocatePediatricECGandselectthedesiredsetting:
Off
On
Thesettingisentered.