76MonitoringCO2(LoFloOption) ExpressionMR400In structions forUse
Applying the Sampling Line to the Patient
Selectthepatientsamplinglinethatisappropriateforthepatientsizeandapplication.Patient
samplinglineswiththeairwayadapterareintendedforusewithbreathingcircuitsand
anesthesiacircuitsthathaveanintegratedairwayadapter.
WARNING
Patient sampling lines are intended for single-patient use only. Do not clean or disinfect these
items. Follow your hospital’s guidelines for appropriate disposal. Reuse of single-use devices
can result in spread of patient infection, degradation of monitoring performance, or
inaccurate measurements.
CAUTIONS
The accuracy of the data is greatly influenced by the proper use and fitting of the patient
sampling line to ensure proper sampling without the introduction of outside air.
Remove the patient sampling line from the CO2 port when not in use.
Toapplythesamplinglinetothepatient
Step Action
1Ensurethatthesamplinglineisclean,dryandundamaged.Replace
thelineifnecessary.
2InsertthesamplinglineconnectorintotheCO2port(seepage74).
Aclickwillbeheardwhenproperlyinserted.
3 Positionthecannulaonthepatient’sface
by
insertingthenasalprongsintothenostrils.
4Passthetubingovertheearsandbehindthehead,
ensuringthepatient’sheadwillnotrestonanypart
ofthecannulawhilethepatientislyingdown.
5Slidethesleevetowardthepatient’sheadtoassurea
goodfitofthe
cannula.
6Ifusingthesamplinglinewiththeairwayadapter,proceedtostep
7;otherwisegotostep8.