User Manual

Table Of Contents
ExpressionMR400InstructionsforUse MonitoringCO2(LoFlo Option)79
Changing the CO2 and CO2 (RESP) Alarm Limits
TheCO2(Et)andCO2(Fi)AlarmLimitsmenucanbeaccessedbytouchingthealarmlimitsettings
intheCO2VSbox.
TochangetheCO2(Et)andCO2(F i)alarmlimitsettings
1 Alarm limit settings, CO2 (Et), CO2 VS box
2High button
3Low button
4CO2 (Et) Alarm Limits menu label (active adjustment shown)
5CO2 (Fi) Alarm Limits menu label
6Enter button
7 Current adjustment
8 Alarm limit, minimum
9 Lower alarm limit setting
10 Upper alarm limit setting
11 Alarm limit, maximum
Step Action
1 Selectthe(Et)CO2alarmlimit
settingsintheCO2VSbox.
TheCO2AlarmLimitsmenuappears.CurrentCO2(Et)settingsare
displayed.
2 SelecttheCO2alarmlimitsmenu,CO2(Et)orCO2(Fi),thatyou
wanttochange.
Theassociatedmenuappears.Currentsettingsaredisplayed.
1
2
7
3456
8 9 10 11