824MonitoringInvasiveBloodPres sure ExpressionMR400InstructionsforUse
Zero Set
ZerosthepressuretransducerforP1(orP2).Thepressuretransducermustbezeroedbeforeuse
andatregularintervalsduringuse.(ZeroPressureChannelwillbedisplayedduringtheprocess
andDonewillbedisplayeduponcompletion.)
Tozerothepressuretransducer
SeeZeroingthePressureTransduceronpage
816.
Set Label
Assignsalabeltothepressurechannelforidentificationofthetransducersite.Thelabelwill
appearintheVSboxanditwillalsodeterminethecolorusedfortheVSbox.
Thefollowingnamesandcolorsareavailable:
Nonedisplayedinwhite.(Default)
ABP(arterialbloodpressure)displayed
inpink.
Note
In the event ABP is chosen as the heart rate source while there are multiple pressures labeled as
ABP, the system will choose P1.
PAP(pulmonaryarterypressure)displayedinyellow.
2Touchthemenuitemtoselect oneofthefollowingP1(orP2)
options:
ZeroSet
SetLabel
Size
HRSource
Grids
GridsSize
Format
Themenuitemappears.Thecurrentsettingishighlighted.
3 Selectthedesiredsettingfromthemenuoptions(exceptGrids,
whichisselectableon
theP1[orP2]menu):
Thesettingisentered.
4Tochangeothersettings,repeatsteps2and3.