User Manual

Table Of Contents
ExpressionMR400InstructionsforUse MonitoringAgentsandGases(AGENTOption)99
WARNINGS
Always test sampling line adapter for a tight connection and proper operation before
attaching to a patient. Over-tightening the sampling line connection may damage the
water trap. Tighten the sampling line connector no more than one half-turn. Over-
tightening this connector can cause failure of the water trap assembly and inaccurate
patient gas measurements.
Inspect water trap and AGENT accessories before use. If the sampling line, connector
or sample port show signs of damage, replace the part immediately or discontinue use
and contact technical support. Never use damaged equipment.
Frequently inspect the patient sampling line and keep it clear of any moving
mechanisms (for example, table wheels) which could cut, pinch, or dislodge the patient
tubing. Avoid kinking of the patient sampling line as leaks, reduced or stopped flow, or
internal venting of sampled gas into damaged tubing will cause inaccurate
measurements.
Do not position the sampling line in any manner that may cause entanglement or
strangulation.
Replace the sampling line if excessive secretions are observed, as inaccurate
measurements could result if the flow is reduced or stopped.
Leakages in the breathing system or sampling system may cause the displayed
AGENT, CO2, O2, N2O values to be too low. Always connect all components securely
and check for leaks according to standard clinical procedures. Displacement of the
nasal cannula or patient airway adapter can cause lower than actual readings.
If AGENT, CO2, O2, N2O values for patients who are not on a breathing circuit
appear extremely low, check whether the patient is breathing through the mouth or
whether one nostril is blocked.
2AllowtheAGSsystemtorunandsampleroomairforatleast1
minute.TheFiO2
readingdisplayedintheGASVSboxshouldbe
approximately21percent.
Ifthereadingremainsoutsidethisrangeformorethan 1 minute
afterfirstcheckingthereading,replacetheoxygensensoras
describedinchapter14.
3 AfterallowingtheAGSsystemtorunforatleast1minute,pinch
or
sealtheinputline(tothewatertrap)for5secondsandverifythat
Occlusionisdisplayed.
Ifthismessagedoesnotappear,checkalltubingconnectionsfor
leakageandretest.
Step Action