ExpressionMR400InstructionsforUse MaintenanceandTroubleshooting145
Theuseroftheproductmustinstitutearoutine checksprogramasdetailedinthetablebelow.
Theuseroftheproductshallmakesurethatallchecksandactionshavebeensatisfactorily
completedbeforeusingtheproduc tforitsintended
purpose.
PlannedTestingProgram
Plannedmaintenancemayonlybecarriedoutbyqualifiedandauthorizedpersonnel,andis
comprehensivelydescribedintheserv icedocumentation.Philipsprovidesafullplanned
maintenanceandrepairserviceonbothacallbasisandacontractbasis. Fulldetailsareavailable
fromyourPhilipsServiceOrganization.
Asummaryofserviceeventsandmaintenance
requirementsappearsinthetablebelow.
Area / Item Frequency Required Action
Accessories Daily Clean and inspect for damage; see page 14-7.
Alarms Daily Confirm proper function; see Testing Alarms on page
14-11.
Module batteries Every 8 hours
of use
Recharge; see page 2-16 for displayed indications, and
refer to the battery charger’s IFU for recharging
instructions.
Wireless module If dropped Inspect and clean (see page 14-9) then test (see page
14-12).
Area / Item
Frequency Required Action
Monitor cleaning Every 6 months Perform external cleaning
System -
AGENT option
Once per year
or as specified
by local laws
Perform calibration.
System - CO2
(LoFlo option)
Once per year
or as specified
by local laws
Perform calibration.
System- NIBP Once per year
or as specified
by local laws
Perform calibration.
System Once per year
or as specified
by local laws,
or after any
type of repair
or service
event
Perform Visual Inspection, and Power On, Verification
and Safety Tests