Cleaning, Disinfecting, and Inspecting MR400 and Wireless Modules
FollowthegeneralguidelinestocleantheMR400cartandthewECGandwSpO2modules.
Alwaysturnoffthecartandthewirelessmodulestoperform cleaning.Never immerseany
portionofthecartorwirelessmodulesinfluidorattempttocleanthedevicesbydirectly
applyingliquidcleaningagents.
Duringthecleaningprocess,inspecttheMR400andthewirelessmodulesfordamaged,looseor
missinghardware;iffound,takecorrectiveactionorcontacttechnicalsupport.
WARNING
Always disconnect the MR400 from AC mains power, and remove the batteries from the
cart and the wireless modules, before performing any cleaning or maintenance. To avoid an
electrical hazard, never immerse any part of the MR400 in any cleaning agent or attempt to
clean it with liquid cleaning agents.
CAUTIONS
Other than those specified in the preceding table, avoid ammonia-, phenol- and acetone-
based cleaners as they will damage the surfaces of the MR400.
Disinfect the MR400 cart and wireless modules as determined by your hospital’s policy to
avoid long term damage to the product.
Do not permit liquid to contact the front or rear of the display panel. Do not permit liquid to
drip into or around the touch screen. Contact technical support if liquid enters any
component.
If the MR400 becomes accidentally wet during use, discontinue operation until all affected
components have been cleaned and permitted to completely dry. Contact technical support if
additional information is required.
Tocleananddisinfectthecartandwirelessmodules
Step Action
1TurnofftheMR400anddisconnectallpower;seepage142.
2 Removethebatteryfromeachwirelessmodule;seepage144.
3Cleanthetouchscreenbywipingitgentlyusingasoftnonwoven
clothwith80%dilutedalcohol
mixture.
4 Removeallvisibledebrisfromthecartandwirelessmodulesusing
soapandwater.
CAUTIONS
Never pour liquid onto the equipment.