124ImportantInformationExpressionMR400InstructionsforUse
Charging Module Batteries
Modulebatteriesmustbechargedforatleast4hoursbeforeinitialuse.Modulebatteriesare
chargedinthePhilipsspecifiedbatterycharger.Refertotheinstructionsprovidedwiththis
batterychargerforinformation.
Installing Batteries in the wECG Module
ThewECGmodulecanacceptuptotwobatteries.Dependinguponthenumberofbatteries
installed,thewECGmoduleprovidedifferentoperationalfeatures:
•Ifonebatteryisinstalled,thenthewECGmodulewillturnonandfunctionnormally—but
beforeitschargeisexhausted,asecondbatterymustbeinstalledinorder
tocontinuethe
ECGstudy.
•Iftwobatteriesareinstalled,thenseamlessoperationispossible—onebatterywillprovide
poweruntilitschargeisexhausted,atwhichtimethewECGmodulewillautomatically
switchtotheremainingbatteryforcontinuedoperation. Aslongassufficientpoweris
providedbythesecondbattery,
continuedoperationispossible.And,anexhausted
batterycanbereplacedatanytimewithoutinterruptiontotheECGstudy,providedthat
sufficientchargeispresentontheremainingbattery.
•IndicatorsidentifythesourcebatterybeingusedbythewECGmodule;seepage29.
•Whenbothbatteriesareremoved,the
wECGmodulewillturnoff.
ToinstallbatteriesinthewECGmodule
Step Action
1HoldthewECGmodulesothatitsbatterybaysareorientedas
shown,withbatterybay1inthelowerpositionandbatterybay2
intheupperposition.
wECG module
Bay 1
Bay 2