ECGNotchFilterappliesanotchfiltertotheECGsignal:
–Off
–50Hz
60Hz(Default)
PressureUnitssetstheunitofmeasureforP1,P2, andNIBPpressurereadings:
–mmHg(Default)
–kPa
GasUnitssetstheunitofmeasurefor
CO2pressurereadings:
mmHg(Default)
–kPa
ECG Tests
AccessestestingfunctionsforECG.
ToopentheECGTestsmenu
PresstheSetupkeyandthenthe
Monitorkey.OntheMonitorSetup
menu,selectService(BioMed). Onthe
Service(BioMed)menu,selectECG
Tests.
Thefollowingmenuitemsare
available:
ECGTestSignalcontrolsawECG
modulegenerated
1mVpeak
topeaksquarewave,where60
BPMwillbedisplayedintheECG
VSboxandECGTestSignalwillbedisplayedinthealarmflagarea(andprintedwhen
outputtotheprinter,ifequipped):
Offdoesnottransmitthetestsignal.(Default)
Ontransmitsthetest
signal.
1ECG Signal Test