User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
4‐26AlarmsExpressionMR400InstructionsforUse
Listing of Alarms
Thissectionlistspatientalarmsalphabetically,andtechnicalalarms(INOPs)arrangedbythe
sourceoftheINOPandthenalphabetically,irrespectiveofpriority.AllalarmsandINOPsare
listedhere;thosewhichcanappearonyourMR400willdependon theinstalledoptions.
SPO2 Saturation % All 1 100 none none
P1 and P2 Systolic mmHg Adult -30 250 -30 250
P1 and P2 Mean mmHg Ped -30 250 -30 250
P1 and P2 Diastolic mmHg Neo -30 250 -30 250
P1 and P2 Pulse Rate BPM All 30 250 30 250
NIBP Systolic mmHg Adult 30 270 30 270
NIBP Systolic mmHg Ped 30 180 30 180
NIBP Systolic mmHg Neo 30 130 30 130
NIBP Mean mmHg Adult 20 255 20 255
NIBP Mean mmHg Ped 20 160 20 160
NIBP Mean mmHg Neo 20 120 20 120
NIBP Diastolic mmHg Adult 10 245 10 245
NIBP Diastolic mmHg Ped 10 150 10 150
NIBP Diastolic mmHg Neo 10 100 10 100
Temperature Temperature °C All 20.0 44.0 20.0 44.0
CO2 (LoFlo option) CO2 (Et) mmHg All 0 90 none 90
CO2 (LoFlo option) CO2 (Fi) mmHg All 0 90 none 90
CO2 (LoFlo option) Resp. Rate RPM All 4 100 none 100
CO2 (AGENT option) CO2 (Et) mmHg All 0 90 none 90
CO2 (AGENT option) CO2 (Fi) mmHg All 0 90 none 90
CO2 (AGENT option) Resp. Rate RPM All 4 100 4 100
AGENT Desflurane (Et & Fi) % All 0 18.0 none 18.0
AGENT Enflurane (Et & Fi) % All 0 5.0 none 5.0
AGENT Halothane (Et & Fi) % All 0 5.0 none 5.0
AGENT Isoflurane (Et & Fi) % All 0 5.0 none 5.0
AGENT Sevoflurane (Et & Fi) % All 0 8.0 none 8.0
AGENT N2O (Et) % All 0 100 none none
AGENT N2O (Fi) % All 0 100 none none
AGENT O2 (Fi) % All 0 100 none none
Bellows Respiration Resp. Rate RPM All 4 150 4 150
Vital Sign or
Parameter
Numeric
Item
Units
Patient
Type
Measurement
Range
Over / Under
Values
Low High Under Over