User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
![](/manual/invivo/3gspo2/user-manual-english/images/img-138.png)
ExpressionMR400InstructionsforUseAlarms4‐29
Violated CO2
respiration rate value
CO2, AGENT The CO2 respiration rate
measurement has violated an
alarm limit setting.
Depending upon RESP VS box configuration:
• Flashing yellow respiration rate numeric
in the CO2 VS box, yellow alarm light,
medium priority alarm tone, or
• Flashing yellow respiration rate numeric
in the RESP VS box, yellow alarm light,
medium priority alarm tone
• And where OVR displayed if the value is
greater than the highest specified
• And where UND if the value is less than
the lowest specified (AGENT option
only)
Missing CO2 respiration
rate data
CO2, AGENT The CO2 respiration rate data,
once present, can no longer be
produced.
Depending upon RESP VS box configuration:
• Flashing yellow dashes (- - -) in the
respiration rate numeric in the CO2 VS
box, yellow alarm light, medium priority
alarm tone, or
• Flashing yellow dashes (- - -) in the
respiration rate numeric in the RESP VS
box, yellow alarm light, medium priority
alarm tone
Violated invasive blood
pressure systolic value
P1 (and P2) The P1 (or P2) invasive blood
pressure systolic measurement has
violated an alarm limit setting.
Flashing yellow systolic numeric in the P1 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified or UND if
the value is less than the lowest specified
Violated invasive blood
pressure mean value
P1 (and P2) The P1 (or P2) invasive blood
pressure mean measurement has
violated an alarm limit setting.
Flashing yellow mean numeric in the P1 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified or UND if
the value is less than the lowest specified
Violated invasive blood
pressure diastolic value
P1 (and P2) The P1 (or P2) invasive blood
pressure diastolic measurement
has violated an alarm limit setting.
Flashing yellow diastolic numeric in the P1 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified or UND if
the value is less than the lowest specified
Missing invasive blood
pressure data
P1 (and P2) The P1 (or P2) invasive blood
pressure data, once present, can
no longer be produced.
Flashing yellow dashes (- - -) in the systolic,
diastolic and mean numerics in the P1 VS box,
yellow alarm light, medium priority alarm tone
Violated primary agent
Et value
AGENT The end-tidal measurement for the
primary agent gas has violated an
alarm limit setting.
Flashing yellow primary agent Et numeric in the
AGENT VS box, yellow alarm light, medium
priority alarm tone, where OVR displayed if
the value is greater than the highest specified
Violated primary agent
Fi value
AGENT The fractional inspired
measurement for the primary
agent gas has violated an alarm
limit setting.
Flashing yellow primary agent Fi numeric in the
AGENT VS box, yellow alarm light, medium
priority alarm tone, where OVR
displayed if
the value is greater than the highest specified
Patient Alarm From Condition Indication