TEMP
Transducer Faulty
P1 (or P2) alarm flag area,
yellow alarm light, medium
priority alarm tone
A electrical transducer malfunction has been detected.
Check the transducer cable connection.
Replace the transducer.
Contact technical support or authorized service personnel if the message persists.
Transducer Not Present
P1 (or P2) alarm flag area,
yellow alarm light, medium
priority alarm tone
An IBP transducer was not found. Ensure that the transducer cables are connected to the
transducer and MR400.
If the message persists, contact technical support or authorized service personnel.
Message, Location,
Indication
What to Do
Cal Error
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
Calibration error. Reconnect the sensor and then retry. Contact technical support or authorized
service personnel if the message persists.
Chk Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature sensor connection is bad, has a sharp bend, or is damaged:
Ensure that the sensor is inserted completely in the temperature port on the patient
connection panel.
Remove any sharp bends in the temperature sensor.
If the message persists after performing the above actions, then the temperature sensor
most likely is damaged and readings cannot be provided. Try a new sensor.
Exp Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe connected to the MR400 is not the proper type. Insert the correct type of
temperature probe into the temperature port.
HW Fail
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
Temperature hardware failure detected. Contact technical support or authorized service
personnel.
No Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe is not properly connected to the MR400. Insert a temperature probe into
the temperature port on the patient connection panel.
Wrong Prb
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe connected to the MR400 is not the proper type. Insert the correct type of
temperature probe into the temperature port.
Message, Location,
Indication
What to Do