User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
ExpressionMR400InstructionsforUseAlarms4‐37
NIBP
Message, Location,
Indication
What to Do
Communication Error
NIBP alarm flag area, blue
alarm light, INOP alarm tone
An internal NIBP error has occurred. Discontinue use of NIBP and contact technical support or
authorized service personnel.
Deflation Timeout
NIBP alarm flag area, red
alarm light, high priority
alarm tone
Cuff deflation has timed out; displayed if the NIBP cuff deflation period is greater than 80 seconds
(neonatal patient type) or is greater than 150 seconds (adult and pediatric patient types):
• Check the patient.
• Check for proper cuff size and placement.
• Check the cuff and hoses for pinching.
If the message persists, contact technical support or authorized service personnel.
Hardware Error
NIBP alarm flag area, blue
alarm light, INOP alarm tone
NIBP hardware failure detected. Discontinue use of NIBP and contact technical support or
authorized service personnel.
Inflation Timeout
NIBP alarm flag area, blue
alarm light, INOP alarm tone
Cuff inflation has timed out:
• Check the patient.
• Check the cuff and hoses for pinching or leaks.
If the message persists, contact technical support or authorized service personnel.
Measurement Failed
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The NIBP measurement has failed:
• Check the patient.
• Check for proper cuff size and placement.
• Check the cuff and hoses for pinching or leaks.
If the message persists, contact technical support or authorized service personnel.
Measurement Timeout
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The NIBP measurement has timed out:
• Check the patient.
• Check for proper cuff size and placement.
• Check the cuff and hoses for pinching or leaks.
Check the cuff condition and placement on the patient.
Module Not Calibrated
NIBP alarm flag area, blue
alarm light, INOP alarm tone
NIBP is not calibrated. Contact technical support or authorized service personnel.
Over Pressure
NIBP alarm flag area, blue
alarm light, INOP alarm tone
The allowed pressure for the type of patient has been exceeded:
• Ensure that the patient is immobilized and not applying pressure to the cuff.
• Check the cuff condition and placement on the patient.
• Make sure that the hose is not pinched.
If the message persists, contact technical support or authorized service personnel.
Pressure Correction
NIBP alarm flag area, blue
alarm light, INOP alarm tone
A pressure correction error has been detected:
• Ensure that the patient is immobilized and not applying pressure to the cuff.
• Check the cuff condition and placement on the patient.
Residual Pressure
NIBP alarm flag area, blue
alarm light, INOP alarm tone
Residual pressure remains in the NIBP system. Disconnect the NIBP hose from the patient
connection panel and then restart the procedure.