User Manual

Table Of Contents
ExpressionMR400InstructionsforUseMonitoringSPO265
ToapplyadisposableneonateSpO2griptothepatient
WAR NIN G
Disposable SpO2 attachments are designed for single patient use and must be disposed after
use. They must not be cleaned and reused. Follow your hospital’s guidelines for appropriate
disposal. Reuse of single-use devices can result in spread of patient infection, degradation of
monitoring performance, or inaccurate measurements.
Step Action
1 Selecttheapplicationsite.ItshouldmatchtheSpO2gripsizeso
thattheattachmentdoesnotfallofforapplyexcessivepressureat
thesite.
2Liftoffthereleaselinersthatprotecttheadhesive.
3 Proceedaccordingtotheapplicationsite:
Footapplication:Alignthehingeontheoutsidefacing
ridge
ofthefoot.Makesuretheattachmentisasfaraspossibleto
wardthesmalltoebutnotoverit.
•Hand/Wristapplication:Alignthehingeontheoutsidefac
ingridgeofthehandorwrist.Youmayhavetoswivelthefi
berheadstoanoptimalposition
toeasetheapplication.
4Withthehingealignedwiththeridgeofthefoot/hand/wrist,press
onesidetotheskinandthenwraptheothersidearoundthelimb
pullingthelongfoampiecegently.
5Pressbothfiberheadsgentlytoattach
theadhesives.
6Securethelong erfoampiecebypressing
itfirmlytothefoam/
adhesiveoftheopposingside.
7Ensurethatthetwo fiberheadsareopposingandhave goodskin
contact.Theanglebetweenthetwofiberheadsshouldbeassmall
aspossible,notexceeding45°.Iftheattachmentopenstoomuch,
reattachortryanothersite.