User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
7‐2MonitoringCO2(LoFloOption) ExpressionMR400In structions forUse
Operation and Use
Whenmonitoringanestheticagentgases,thetypicaloperationsandpossibleconditionsthatcan
arisemayresultinpotentialmessagesrequiringyourattention.Seepage4‐26foramessage
listingandsuggestedactions.
Warm-Up Period
Inordertoachieveaccurateidentificationsandmeasurements,theLoFlosystemrequiresa
warm‐upperiodtothermallystabilize.Thiswarm‐upperiodbeginswhentheCO2parameteris
activated.Uponactivation,CO2WarmingUpwillbedisplayeduntiltheLoFlosystembecomes
fullyoperational(about2minutes).
Zero Reference Adjustment
WARNING
During Zero calibration the system pulls ambient air through the zero intake port on the
cart The calibration system assumes that the ambient air will contain normal trace amounts
of CO2. If the system is placed in an unventilated area that allows CO2 (from the waste gas
port on the rear panel, if not connected to a gas scavenging system) to accumulate, the result
could be inaccurate CO2 zeroing and resulting inaccurate patient readings. Always place the
cart in a well ventilated area.
TheLoFlosystemwilloccasionall yperformazeroreferenceadjustment(ZeroCa l )toensurethe
accuracyofthe displayedgasconcentrations.PerformingCO2Zerowillbedisplayedduringazero
referenceadjustment;allowtheprocesstocomplete.Thema ximumtimerequiredforcalibrationis
approximately40seconds.ZeroCal
isnotrequiredwhenswitchingsamplinglines.
Undercertainconditions,ZeroCalwi llnot beallowed:
•Iflessthan20secondshavepassedsincedetectionofthelastbreath;
•IftheCO2temperatureisunstable;or
•IfthesamplinglineisdisconnectedfromtheCO2port.
Toperformamanualzeroreferenceadjustment
Step Action
1
IfCO2WarmingUpisdisplayed,waitasthesystemthermally
stabilizes.Whenthemessageiscleared,proceed.