User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
![](/manual/invivo/3gspo2/user-manual-english/images/img-208.png)
7‐8MonitoringCO2(LoFloOption) ExpressionMR400In structions forUse
alsoprovidedinthisareaofthescreen.(CO2[RESP]informationcanbedisplayedintheCO2VS
boxorintheRESPVSbox,asdetailedbelow.)
Item Name Definition
1 CO2 VS waveform Is the detected CO2 waveform (Trace D)
Note
To change the waveform speed, see Resp Speed on page 3-25.
2 Alarm flag area Displays CO2 alarm flags when detected; see page 4-31.
3
CO2
VS box label Indicates the CO2 vital sign parameter, and accesses the CO2
menu
4 Unit of measure Indicates that the gas measurement numeric values are given in mmHg
(millimeters of mercury) or kPa (kilopascals); see page 7-11.
5 FiCO2 numeric Is the patient’s detected fractional inspired CO2 measurement
6 EtCO2 numeric Is the patient’s detected end-tidal CO2 measurement
7 EtCO2 upper alarm limit
Is the upper limit setting for the end-tidal CO2 alarm, and accesses the CO2
(Et) Alarm Limits menu
8 EtCO2 lower alarm limit
Is the lower limit setting for the end-tidal CO2 alarm, and accesses the CO2
(Et) Alarm Limits menu
9 Respiration rate upper alarm
limit
Is the upper limit setting for CO2-derived respiration rate alarm, and accesses
the CO2 (RESP) Alarm Limits menu
10 Respiration rate lower alarm
limit
Is the lower limit setting for CO2-derived respiration rate alarm, and accesses
the CO2 (RESP) Alarm Limits menu
11 Respiration rate numeric Is the patient’s detected respiration rate measurement, as derived from CO2
12 Unit of measure
Indicates that the respiration rate numeric is given in RPM (respirations per
minute)
When Source is set to BEL in the RESP menu (see page 10-5), the CO2 VS box will also contain CO2-derived respiration rate
elements, as indicated by the shaded rows and illustration above; otherwise, this information will be displayed in the RESP VS box
(see page 7-10).
7
1
3
4
5
6
11
8
9
10
2
12