A10Specifications ExpressionMR400InstructionsforUse
When“HR”isderivedfromSPO2
Lower
Upper
Off,or30to250BPM
60to250BPM,oroff
Note
Measurement validation: SPO2 accuracy validated in human studies against arterial blood sample reference measured with
a CO-oximeter. In a controlled desaturation study, healthy adult volunteers with saturation levels between 70–100 percent
SPO2 were studied. The population characteristics for those studies were:
about 50% female and 50% male subjects
19 – 27 years of age
light to black skin tones
Reference method for the computation of pulse rate accuracy made using an electronic pulse simulator. (A functional
tester cannot be used for accuracy assessment of a pulse oximeter monitor; however, it can demonstrate that a pulse
oximeter monitor reproduces a calibration curve that has been independently demonstrated to fulfill a particular
accuracy specification.)
SPO2 measurements are statistically distributed; therefore, in accordance to 80601-2-61:2011, it is possible that only
two-thirds of the measurements will fall within ± 3 percent of the value measured by the CO-Oximeter.
CO2 (Optional LoFlo)
Sidestream,nondispersiveinfraredabsorptiontechnique,includesmultiplefiltrationsystemandmicroprocessor
logiccontrolofsamplehandlingandcalibration.MethodfordeterminingendtidalCO2measurement:Measures
peakoftheendtidalCO2waveformevery20seconds.
Output CO2waveform,EtCO2
andFiCO2
measurementnumericvalues,
andrespirationrate
InitializationTime Waveformdisplayedinlessthan20seconds,atanambient
temperatureof25°C(77°F);fullspecificationsattainedwithin2
minutes
ZeroCalibrationInterval Automaticoruserrequested
CO2UnitofMeasure Millimetersofmercury(mmHg)orkilopascals*(kPa)
CO2Resolution 1mmHg(0.1kPa)
FlowRate 50mL/minute±10
mL/minute
DataSampleRate 100Hz
EndtidalCO2(EtCO2)MeasurementRange
(InwhichtheCO2accuracyspecificationismet)
0to76mmHg(0to10.1kPa)forrespiration ratesrangingfrom4
to60breathsperminute,inclusive
Pulse Oximeter