User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
A‐12Specifications ExpressionMR400InstructionsforUse
Compensations
(AutomaticCO2ambientpressure
compensation400–800mmHg[53.3–106.6
kPa] )
Forend‐tidalO
2
balancegas(N
2
,N
2
O,O,He)andanesthetic
agents
B
Usesgascompensationinformationtocorrecttherawcarbon
dioxidevalue
AnestheticAgentEffects
(MACLevels)
AnestheticAgentSensitivity
A
(uncompensated):Accuracy
specificationwillbemaintainedforhalogenatedanesthetic
agentspresentatacceptedMAC(MinimumAlveolar
Concentration)clinicallevels.
AnestheticAgentSensitivity(compensated):Testingatagent
levelsdefinedbyacceptedregulatorystandards(60601‐2‐55)
Cross‐sensitivityCompensationError
(Additionalworstcaseerrorwhen
compensationforO
2
,N
2
O,anestheticagents,
orheliumiscorrectlyselectedfortheactual
fractionalgasconstituentspresent.)
0to40mmHg:±1mmHgadditionalerror
(0to5.3kPa:±0.1 kPaadditionalerror)
41to70mmHg:±2.5mmHgadditionalerror
(5.5to9.3kPa:±0.3kPaadditionalerror)
71to
100mmHg:±4mmHgadditionalerror
(9.5to13.3kPa:±0.5kPaadditionalerror)
101to150mmHg:±5mmHgadditionalerror
(13.5to20kPa:±0.6kPaadditionalerror)
Quantitativeeffectsofgassamplehumidityor
condensate**:
0to40mmHg:±2mmHg
(0to5.3kPa:±
0.2kPa)
41to70mmHg:±5percent
(5.5to9.3kPa:±5percent)
71to100mmHg:±8percent
(9.5to13.3kPa:±8percent)
101to150mmHg:±10percent
(13.5to20kPa:±10percent)
**Withappropriatecompensationsapplied
Note
There are no known adverse effects on stated performance due to cyclical pressure of up to 10 kPa (100 cmH2O).
CalibrationInterval Calibrationverificationmustbeperformedat1yearintervals.
CO2 (Optional LoFlo)