User Manual

Table Of Contents
ExpressionMR400InstructionsforUse SpecificationsA17
SystemResponseandRiseTimes
(Asmeasuredfrompatientgasinputofthe
completepneumaticcircuit,includingtubing,
from1090percentofmeasuredlevels)
Cannula,Adult Halothane—
Systemresponse:11.56seconds
Risetime:5.77seconds
Enflurane—
Systemresponse:7.55seconds
Risetime:1.75seconds
Isoflurane—
Systemresponse:6.71seconds
Risetime:0.88seconds
Sevoflurane—
Systemresponse:6.45seconds
Risetime:0.62seconds
Desflurane—
Systemresponse:6.63seconds
Risetime:0.57seconds
Oxygen—
Systemresponse:6.99seconds
Risetime:1.02seconds
Nitrousoxide—
Systemresponse:6.28seconds
Risetime:0.25seconds
CO2—
Systemresponse:6.62seconds
Risetime:0.61seconds
AGENT (Optional)