Using SPO2 Gating
ToreceiveSPO2gatingsignals
Step Action
1CheckthebatteryindicatoronthewSpO2moduletoensurethat
enoughchargeexists:
Greenbatteryindicator=Chargesufficient;proceedtostep5.
•Redbatteryindicator=Chargelow;proceedtostep4.
Seepage211fordetails.(Also,youcanreferencethestatus
information
pane;seepage216.)
2InsertachargedmodulebatteryintothewSpO2moduleandthen
recheckthebatteryindicatortoensureasu fficientchargebefore
proceeding;seepage126.
3CheckthenetworkchannelindicatoronthewSpO2moduleto
ensurecommunicationisestablishedwiththeMR400:
Steady=Goodcommunication;
proceedtostep7.
Flashing=Nocommunication;proceedtostep6.
Seepage211fordetails.(Also,youcanreferencethestatus
informationpane;seepage216.)
4Ensurethatthecorrectgatingcableisinstalledbetweenthe
MR400andtheMRsystem.
5EnsurethatthewSpO2moduleiswithin
9.1m(30feet)ofthe
MR400,inthesameMRIroomorinthesameshieldedroom,andis
settothesamewirelessnetworkchannelusedbytheMR400;see
page129.
6
EnsurethattheSPO2attachmentisproperlyattached;seeApply
ingtheSPO2Attachmenttothe
Patientonpage63.
7 SelecttheSPO2VSbox(seepage 68).
TheSPO2menuappears.Currentsettingsaredisplayed.
8 SelectGatingSource.
TheGatingSourcemenuappears;seepage515.
9 SelectPulse.
Thesettingisapplied.