989803173791 Rev 0.6Chapter3:PreparationforUse37
Chapter 3: Preparation for Use
TheEssentialMRIPatientMonitor(Model865353)providestheflexibilityneededtoperform
standardSpO
2
monitoring,whileallowingyoutocustomize operationstofityourneeds.
Warning
WAR N I N G
Alwaysverifypropercommunicationsbetweenthemoduleandmonitorpriortopatientuse.
CAUTIONS
•Avoidtheuseofcellularphonesorotherradio‐frequencytransmittersintheproximityof
anoperatingsystem.
•AminorbutnoticeabledegradationinthewirelessSpO
2
radiocommunicationswilloccur
inthepresenceofhigh‐poweredradios.
Using the Monitor
ObserveallwarningsandcautionswhenusingtheEssentialMRIPatientMonitor(Model
865353).
Warning
WAR N I N G
KeeptheEssentialdisplayoutoftheMRsystemboreandavoidcontactwiththepatient’sbare
skinwhiletheMRIscanisrunning.Thedeviceiselectronicandsusceptibletoexcessiveheating
fromtheMRIscanRFonlyifplacedinsidetheboreoftheMRsystemduringthatscan.
Failure
todosomayresultinpatientinjury.
Warning
WAR N I N G
DonotuseortakethepoweradapterinsidetheMRsystemroom.Thedeviceismagneticand
willbepulledintotheMRsystem.Thedeviceisintendedforusewiththemainbatteryandthe
modulebatteryonlywhenoutsidetheMRsystemroom.