User's Manual

Table Of Contents
989803173791 Rev 0.6Chapter3:PreparationforUse47
HRAlarmLimitsbuttons:Indicatesthecurrentsettingandallowsyoutosetthelimitsfor
theHR(heartrate)alarm,adjustedbytouchingthebutton(HighorLow)followedbythe
incrementordecrementbutton.
•Decrementbutton:AllowsyoutodecreaseanAlarmLimits
setting,wheretouchingitonce
decreasesthecountbyoneandholdingitdecreasesthecountcontinuously.
•Incrementbutton:AllowsyoutoincreaseanAlarmLimitssetting,wheretouchingitonce
increasesthecountbyoneandholdingitincreasesthecountcontinuously.
DefaultLimitsbutton:Allowsyoutoautomaticallychange
thelowandhighalarmlimits
forbothparameterstothedefaultsettings(seethetablesbelow).
DisplayLimitsbutton:Allowsyoutocontrolthepresenceofthehighandlowalarmlimit
indicatorsforSpO
2
andHR,whereacheckedboxdisplaystheindicators.
SpO2vitalsignnumeric:IndicatestheSpO
2
vitalsignofthepatient(givenasapercentage)
andallowsyoutochangetotheSpO
2
alarmlimits,wheretouchingit openstheAlarm
Limitsmenu.
HRvitalsignnumeric:Indicatestheheartratevitalsignofthepatient(giveninbeatsper
minute)andallowsyoutochangetotheheartratealarmlimits,wheretouchingitopens
theAlarmLimitsmenu.
SpO2highalarmlimit
indicator:Indicatesandallowsdirectchangestothehighlimit
settingfortheSpO2alarm,wheretouchingitopenstheAlarmLimitsmenu.
SpO2lowalarmlimitindicator:Indicatesandallowsdirectchangestothelowlimitsetting
fortheSpO2alarm,wheretouchingitopenstheAlarmLimitsmenu.
HRhighalarmlimitindicator:Indicatesandallowsdirectchangestothehighlimitsetting
fortheheartratealarm,wheretouchingitopenstheAlarmLimitsmenu.
HRlowalarmlimitindicator:Indicatesandallowsdirectchangestothelowlimitsetting
fortheheartratealarm,wheretouchingit
openstheAlarmLimitsmenu.
NeonatalAlarmLimits
VitalSign
Parameter
Unit
LowLimit HighLimit
MIN MAX Default MIN MAX Default
SpO
2
Percent 50 99 90 70 99 Off
HR BPM 30 249 90 60 249 210
PediatricAlarmLimits
VitalSign
Parameter
Unit
LowLimit HighLimit
MIN MAX Default MIN MAX Default
SpO
2
Percent 50 99 90 70 99 Off
HR BPM 30 249 75 60 249 160