48Chapter3:PreparationforUse989803173791 Rev 0.6
Tochangeanalarmsetting,followthestepsbelow:
AdultAlarmLimits
VitalSign
Parameter
Unit
LowLimit HighLimit
MIN MAX Default MIN MAX Default
SpO
2
Percent 50 99 85 70 99 Off
HR BPM 30 249 45 60 249 160
Step Action
1
Ifachangeinthepatientcategoryisneeded,touchthePatient
categorybuttonthenmakethatchange;otherwise,proceedto
Step2.
2TouchtheAlarmsbutton .
3TouchthehighorthelowAlarmLimitsbuttonfortheparameter.
(Thetouchtonewillsoundandthebuttonwillbehighlighted.)
TheexamplebelowfollowsaSpO
2
highalarmlimitchange.
4Touchthedecrementortheincrementbuttonuntilthedesired
settingisreached,asdisplayedinthehighlightedAlarmLimits
button.
Forexample,thedecrementbuttonwastouchedtwice,resultingin
ahighalarmlimitof98.
5 RepeatSteps2and3toadjustadditionalalarmlimits.