989803173791 Rev 0.6Chapter5:Workflow 67
WorkflowwhenusingtheEssentialwithanExpression(oraPrecess‐Blueversion)systemand
oneWSpO
2
module:
•Apatient,anMRinductionroom,anMRsystemroomwithanExpression(oraPrecess)
andanMRrecoveryroom.
–TheEssential,theExpressionandtheWSpO2modulearesettonetworkA(orifusing
aPrecess‐Blueversion,network1).
5Setthepatientcategoryandverifypulsatile
flow.
6TransportthepatientwiththeEssentialintotheMRsystemroom.
7Scanthepatient,monitoringSpO
2
withtheEssentialthroughthe
controlroomwindow.
8TransportthepatientwiththeEssentialandtheWSpO
2
module
intotheMRrecoveryroom.
9 TransitionthepatientintotheMRrecoveryroomtobedside
monitoring,ifneeded,thentaketheEssentialandtheWSpO
2
moduleintotheMRinductionroomforthenextpatient.
Step Action
1PrepthepatientintheMRinductionroom.
2TurnontheEssentialandtheWSpO
2
module.Botharesetto
networkA(or1ifusingaPrecess).
3Verifysufficientpowerandgoodcommunications.
4ApplytheSpO
2
sensortothepatient.
5Setthepatientcategoryandverifypulsatileflow.
6TransportthepatientwiththeEssentialintotheMRsystemroom
forscanning,wheretheExpressionissettonetworkA(orifusinga
Precess,network1).