User's Manual

Table Of Contents
70Chapter5:Workflow989803173791 Rev 0.6
WorkflowwhenusingtheEssentialwithaPrecess(Blueversion)system,twoWSpO
2
modules
andtwoECGmodules:
•Twopatientssimultaneously,anMRinductionroom,anMRroomwithaPrecessandan
MRrecoveryroom.
–ThePrecess,aWSpO
2
moduleandanECGmodulearesettonetwork1.
TheEssential,aWSpO
2
moduleandanECGmodulearesettonetwork2.
Step Action
1One(current)patient isintheMRsystemroom,whereaPrecess,a
WSpO
2
moduleandaECGmodulearesettonetwork1.
2Prepthesecond(new)patientintheMRinductionroom.
3TurnontheEssential,theWSpO
2
moduleandtheECGmodule.All
aresettonetwork 2.
4Verifysufficientpowerandgoodcommunications.(TheECG
modulecannotnotbecheckedatthispoint.)
5ApplytheSpO
2
sensorandtheECGleadstothenewpatient.
6Setthepatientcategoryandverifypulsatileflow.
7Whenthescanofthecurrentpatienthasfinished,transportthe
newpatientwiththeEssential,theWSpO
2
moduleandtheECG
moduleintotheMRsystemroom.
8IntheMRsystemroom,onthePrecessdisplaypanel,pressthe
MONITORSETUPkey.TurnthecontrolknobtoNETWORK,pressthe
knobandselectNetwork2.
NOTE
PresstheAlarmsSilencekeytotemporarilydisablethealarm
functions.
9IntheMRsystemroom,connectthenewpatienttotheadditional
vitalmeasurementsonthePrecess.Verifysufficientpowerand
goodcommunicationsforthemodules.