74Chapter6:MaintenanceandRepair989803173791 Rev 0.6
CAUTIONS
•Avoidammonia based,phenolbased,andacetonebasedcleaners.Theywilldamage the
systemsurfaces.
•Ifthesystembecomesaccidentallywetduringuse,discontinueoperationuntilallaffected
componentshavebeencleanedandpermittedtodrycompletely.ContactInvivoTechnical
Supportifadditionalinformationisrequired.
Cleaning the Accessories
Anyreusablepatientaccessoriesmustbecleanedaftereachuse.Disposable patientaccessories
mustbediscardedandreplacedwithnewitems.Theaccessoriescannotbesterilized.
WAR N I N G
Singleusedevices,asindicatedonthedevicepackaging,shouldbedisposedofafter useand
mustneverbereused.
CAUTION
Neverimmerseanaccessoryinanyc leaningfluid.
Tocleanreusableaccessories(suchasSpO
2
sensorsandgrips),completethefollowingsteps:
CAUTION
Disinfecttheaccessoryasdeterminedbyyourfacility’spolicy.
Step Action
1 Removetheaccessoryfromuse.
2 Removedirtanddustfromtheaccessoryusingalintfreecloth,
moistenedwithwarmwater(40CFmaximum)gentlywiping
allsurfacestobecleanedbriefly(30secondsto1minute)as
neededtoensure
propercleansing.Stainscanberemovedfrom
theaccessorybyscrubbingbrisklywiththemoistenedcloth.
3Inspecttheaccessoryforanycracks,holes,tears,cuts,etc.that
couldaffectoperationandreplaceasnecessary.
4Ifdisinfectionisrequired,useonlytherecommendedliquidsurface
disinfectants,unlessotherwisespecifiedintheaccessory’s
instructionsforuse.Recommendedsurfacedisinfectantsinclude
dilutesolutionsofanyofthefollowing:
•CaviWipes
Alcohol(70%)
AntibacterialSoap(0.1%Triclosan)