User Manual

Table Of Contents
ExpressionMR400InstructionsforUseGettingStarted317
Toeditsavedusersettings
Todeletesavedusersettings
7Ifdesired,presstheSettoDefaultbutton tosavethecurrentfileas
thedefaultsetupforuseatpowerup.
Warning
WAR NIN G
If you choose to boot the device from a user settings
file, confirm that alarm presets are appropriate for the
patient prior to monitoring. Failure to do so may cause
false or missed alarm conditions.
8PresstheSave&Closebuttontosavethecurrentsetupandexit
themenu.
Awarningboxwillappear,promptingyoutoacceptthechanges:
presstheAcceptbuttontosavethecurrentsettings,orpressthe
Cancelbuttontorejectthechanges.
Step Action
1ConfiguretheMR400forthedesired
settings.
2OntheMonitorSetupmenu,selectEditUserSettings.
3 Selectthefilenamethatyouwanttochange.
4PresstheCaptureSettingsbuttontochangetheusersetuptothe
currentsettings.
5PresstheSave&Closebuttontosavethecurrentsetup.
Awarningboxwillappear,promptingyouto
acceptthechanges—
pressAccepttosavethecurrentsettingsorpressCanceltoreject
thechanges.
Step Action
1OntheMonitorSetupmenu,selectEditUserSettings.
2 Selectthefilenamethatyouwantdelete.
Step Action