AGENTallowsanestheticagentandgasmonitoring(butdoesnothaveanassociated
waveform),andCO2and CO2derivedrespirationratemonitoring:
OffturnsofftheAGENTparameter.(Default)
OnturnsontheAG ENTparameter.
NOTE
When AGENT is turned on, CO2 will also be activated, including GAS; however, if AGENT is
then turned Off, CO2 will remain active.
Tocontrolparameters
Sound Adjust
Controlsalarm,heartrateandtouchscreentones,andthevolumesettingsforthesounds
generatedbytheMR400.
Warning
WAR NIN G
The alarm sound can be turned off, as indicated by the symbol. Always ensure that the
alarm sound setting is appropriate for the monitoring environment and for each patient. The
alarm sound volume is adjustable for suitability to various clinical environments. When you
use the MR400, always ensure that the alarm sound can be heard above the ambient noise
level; otherwise, treatment of the patient could be delayed.
ToopentheSoundAdjustmenu
Step Action
1PresstheSetupkeyandthentheMonitorkey.
TheMonitorSetupmenuappears.Currentsettingsaredisplayed.
2OntheMonitorSetupmenu,selectParameters.
TheParametersmenuappears.Currentsettingsaredisplayed.
3Locatetheparameterthatyouwanttocontrolthenselectthe
desiredsetting:
Off
On
Thesettingisentered.Tochangeothersettings,repeatstep3.