ExpressionMR400InstructionsforUseGettingStarted333
NOTE
ECG Test Signal is unavailable when Filter Mode > Advanced 2 is selected; see 5-31.
TocontroltheECGTestSignal
NIBP Tests
AccessescalibrationandtestingfunctionsforNIBP.
NOTE
If Module Not Calibrated is displayed, then all options in the NIBP Tests menu will be locked
until NIBP is successfully calibrated.
ToopentheNIBPTestsmenu
Step Action
1EnsurethatthewECGmoduleiscommunicatingwiththeMR400.
2PresstheSetupkeyandthentheMonitorkey.
TheMonitorSetupmenuappears.Currentsettingsaredisplayed.
3OntheMonitorSetupmenu,selectService(BioMed).
TheService(BioMed)menuappears.Currentsettingsare
displayed.
4LocateECGTestSignalandselectthedesiredsetting:
Off
On
Thesettingisentered.