User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
ExpressionMR400InstructionsforUseAlarms4‐3
AlarmflagsassociatedwithavitalsignaredisplayedalongsidetheVSboxofthatparameter(see
illustrationbelow),whilealarmflagsassociatedwiththesystemaredisplayedinthesystem
messagearea(middleandtopcenterofthetouchscreen,see
page2‐21).Whenmultiplealarm
flagsarepresent,theywillbestackedinacolumn.
Flashing Numeric
Duringanpatientalarmcondition,thevitalsignnumericwillchangecolorandflashtoindicate
thesource,typeandpriorityofthealarm:
•Highprioritypatientalarm—rednumeric,flashingat1.5Hzwitha50%dutycycle.
•Mediumprioritypatientalarm—yellownumeric,flashingat0.75Hzwitha50%duty
cycle.
Whiletheviolationcontinues,thenumericoftheviolatedparameterwillflashinprioritycolorof
thedetectedalarm.
Alarm Light
Duringanalarmcondition,thealarmlight(see
inset)canilluminatetoprovidea360degree,
visualindicationofthealarmpriority, asdetailed
inthetablebelow.
Multiplecolorscanbeilluminatedwhen
multiplealarmconditionsexist.Thealarmlight
isamenu‐controlledfeature;seepage4‐20for
settingdetails.
1 Alarm flag (an INOP in this example)
1
AlarmPriority LightColor Indication
High(patientalarm) Red Flashing,1.5Hzwitha50%dutycycle
Medium(patientalarm) Yellow Flashing,0.75Hzwitha 50%dutycycle
INOP(statusortechnicalalarm) Blue Steady
Alarm light, red or yellow
Alarm light, blue
Speaker
Display panel