428AlarmsExpressionMR400InstructionsforUse
Missing heart rate data ECG, SPO2, P1
(or P2)
The heart rate data, once present,
can no longer be produced.
Flashing yellow dashes (- - -) in the heart rate
numeric in the ECG VS box and in the SPO2
VS box, yellow alarm light, medium priority
alarm tone
Over maximum heart
rate value
ECG, SPO2, P1
(or P2)
The heart rate measurement has
violated the upper parameter
range.
Flashing red heart rate numeric alternating
between OVR and the heart rate value in the
ECG VS box and in the SPO2 VS box, red alarm
light, high priority alarm tone
Under minimum heart
rate value
ECG, SPO2, P1
(or P2)
The heart rate measurement has
violated the lower parameter
range.
Flashing red heart rate numeric alternating
between UND and the heart rate value in the
ECG VS box and in the SPO2 VS box, red alarm
light, high priority alarm tone
Desat SPO2 SPO2 has detected a desaturation
event.
Red flag in ECG alarm flag area, flashing red
SpO2 numeric, red alarm light, high priority
alarm tone
Violated arterial oxygen
saturation value
SPO2 The arterial oxygen saturation
measurement has violated an
alarm limit setting.
Flashing yellow SpO2 numeric in the SPO2 VS
box, yellow alarm light, medium priority alarm
tone
Missing arterial oxygen
saturation data
SPO2 The arterial oxygen saturation
data, once present, can no longer
be produced.
Flashing yellow dashes (- - -) in the SpO2
numeric in the SPO2 VS box, yellow alarm
light, medium priority alarm tone
Apnea CO2, AGENT The apnea time delay setting has
been exceeded.
Red flag in the CO2 or RESP alarm flag area,
Flashing red respiration rate numeric, red
alarm light, high priority alarm tone
Note
Apnea alarm flag appears in either the CO2 or
the RESP alarm flag area, based on the RESP
source.
Violated EtCO2 value CO2, AGENT The end-tidal CO2 measurement
has violated an alarm limit setting.
Flashing yellow EtCO2 numeric in the CO2 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified
Missing EtCO2 data CO2, AGENT The end-tidal CO2 data, once
present, can no longer be
produced.
Flashing yellow dashes (- - -) in the EtCO2
numeric in the CO2 VS box, yellow alarm light,
medium priority alarm tone
Violated FiCO2 value CO2, AGENT The fractional inspired CO2
measurement has violated an
alarm limit setting.
Flashing yellow FiCO2 numeric in the CO2 VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified
Missing FiCO2 data CO2, AGENT The fractional inspired CO2 data,
once present, can no longer be
produced.
Flashing yellow dashes (- - -) in the FiCO2
numeric in the CO2 VS box, yellow alarm light,
medium priority alarm tone
Patient Alarm From Condition Indication