User Manual

Table Of Contents
ExpressionMR400InstructionsforUseAlarms435
P1 (or P2)
Multiple Agents
AGENT alarm flag area, blue
alarm light, INOP alarm tone
More than one anesthetic agent gas was detected in a given breath phase, with a total MAC of the
detected mix is less than 3 MAC.
Note
If multiple agents have been detected with a total MAC of the detected mix greater than 3 MAC then
this message will be accompanied by a yellow alarm light and medium priority alarm tone.
O2 Sensor Not Present
AGENT alarm flag area, blue
alarm light, INOP alarm tone
Possible hardware failure associated with the O2 sensor. Ensure that the O2 sensor is secure in the
receptacle; see page 14-13.
If the message persists, contact technical support or authorized service personnel.
Occlusion
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Occluded sample line detected at start up:
Check the sampling line for obstructions and replace it if necessary.
Check the fluid level in the water trap and replace it if necessary.
If the message persists, contact technical support or authorized service personnel.
O2 Sensor Fail
CO2 alarm flag area, blue
alarm light, INOP alarm
tone
O2 sensor has failed or expired. Replace the O2 sensor; see page 14-13.
If the message persists, contact technical support or authorized service personnel.
Occlusion at Start
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Occluded sample line detected at start up:
Check the sampling line for obstructions and replace it if necessary.
Check the fluid level in the water trap and replace it if necessary.
If the message persists, contact technical support or authorized service personnel.
Performing CO2 Zero
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Displayed while zeroing CO2. Allow the process to complete.
Persistent CO2 Occlusion
CO2 alarm flag area, blue
alarm light, INOP alarm tone
Reduced CO2 flow has been detected for over 2 minutes. Check for pinches or obstructions in the
sampling line. Clear any pinch, or replace the accessory if necessary.
If the message persists, contact technical support or authorized service personnel.
Message, Location,
Indication
What to Do
Catheter Disconnected
P1 (or P2) alarm flag area,
yellow alarm light, medium
priority alarm tone
The catheter cannot be detected. Check all catheter connections to and from the transducer.
Hardware Error
P1 (or P2) alarm flag area,
blue alarm light, INOP alarm
tone
A hardware error has been detected in the IBP system. Contact technical support or authorized
service personnel.
Message, Location,
Indication
What to Do