User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
4‐36AlarmsExpressionMR400InstructionsforUse
TEMP
Transducer Faulty
P1 (or P2) alarm flag area,
yellow alarm light, medium
priority alarm tone
A electrical transducer malfunction has been detected.
• Check the transducer cable connection.
• Replace the transducer.
• Contact technical support or authorized service personnel if the message persists.
Transducer Not Present
P1 (or P2) alarm flag area,
yellow alarm light, medium
priority alarm tone
An IBP transducer was not found. Ensure that the transducer cables are connected to the
transducer and MR400.
If the message persists, contact technical support or authorized service personnel.
Message, Location,
Indication
What to Do
Cal Error
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
Calibration error. Reconnect the sensor and then retry. Contact technical support or authorized
service personnel if the message persists.
Chk Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature sensor connection is bad, has a sharp bend, or is damaged:
• Ensure that the sensor is inserted completely in the temperature port on the patient
connection panel.
• Remove any sharp bends in the temperature sensor.
• If the message persists after performing the above actions, then the temperature sensor
most likely is damaged and readings cannot be provided. Try a new sensor.
Exp Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe connected to the MR400 is not the proper type. Insert the correct type of
temperature probe into the temperature port.
HW Fail
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
Temperature hardware failure detected. Contact technical support or authorized service
personnel.
No Probe
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe is not properly connected to the MR400. Insert a temperature probe into
the temperature port on the patient connection panel.
Wrong Prb
TEMP alarm flag area, blue
alarm light, INOP alarm
tone
The temperature probe connected to the MR400 is not the proper type. Insert the correct type of
temperature probe into the temperature port.
Message, Location,
Indication
What to Do