822MonitoringInvasiveBloodPres sure ExpressionMR400InstructionsforUse
Changing the Unit of Measure
Note
When using an IP5 and Pressure Units is changed, the displayed formatting of the value and
placement of the decimal point is changed immediately. However, it can take up to 2 seconds for
the measurement numeric values to reflect the new unit of measure. Do not print or perform data
captures during this period.
Tochangetheunit ofmeasure
P1 (and P2) Menu
TheP1andP2menusallowyoutocontrolinvasivebloodpressuretraces,functionsandsettings.
Eachmenucontainsidenticaloptionsforthecontroloftherespectiveinvasivebloodpressure
channel,P1orP2.
Note
The operation and menus for P1 and P2 are identical.
Step Action
1PresstheSetupkeyandthentheMonitorkey.
TheMonitorSetupmenuappears.Currentsettingsaredisplayed.
2OntheMonitorSetupmenu,selectService(BioMed).
TheService(BioMed)menuappears.Currentsettingsare
displayed.
3OntheService(BioMed)menu,selectSystemConfig.
TheSystemConfigmenuappears.Currentsettingsare
displayed.
4OntheSystemConfigmenu,selectPressureUnits.
ThePressureUnitsmenuappears.Thecurrentsettingis
highlighted.
5 Selectthedesiredsetting:
mmHg
kPa
Thesettingisentered.