ExpressionMR400InstructionsforUse MaintenanceandTroubleshooting147
WARNING
Do not use flammable or potentially explosive disinfecting sprays. Such sprays create vapors,
which can ignite, causing fatal or other serious personal injury.
CAUTION
Disinfecting a medical product room by means of sprays is not recommended, since the vapor
could penetrate the product, causing electrical short-circuits, metal corrosion or other damage to
the product.
Cleaning, Disinfecting, and Inspecting the Accessories
Anyreusablepatientaccessory(suchasECGcables,SPO2
attachmentsandprobes,temperature
sensors,etcetera)mustbecleanedanddisinfectedbeforeinitialuseandaft ereachuseto
protectpatientsandpersonnelfr omavarietyofpathogens.Usesoapandwater,andasuggested
disinfectantandmethod,tocleananddisinfect theaccessories.Thewarrantydoesnotcover
damage
causedbyunapprovedsubstancesormethods.
Duringthecleaningprocess,inspecttheaccessoryfordamage.Theaccessoriesareexposedto
potentiallydamagingsituationsduringuseandcleaning.Beforeeachuse,carefullyinspectthe
accessoriesforthefollowingsignsofdamage:
•Cracks,holes,tears,gouges,cuts,etcetera.
•Cracksorother
signsofdamagetotheconnector,includingbentordamagedpins.
Disposableaccessoriesmustbediscardedandreplacedwithnewitems.
WARNING
Cracks, tears, cuts and gouges interfere with standard cleaning procedures and therefore
pose a potential risk to patients and personnel. If you see any sign of damage to any
accessory, immediately discontinue use.
WARNING
Single-use devices, as indicated on the device packaging, should be disposed of after use and
must never be reused.Reuse of single-use devices can result in spread of patient infection,
degradation of monitoring performance, or inaccurate measurements.
CAUTIONS
Never immerse an accessory in any cleaning fluid.
Do not autoclave any part of the equipment. Disinfect the accessory as determined by your
facility’s policy.