User Manual

Table Of Contents
ExpressionMR400InstructionsforUseImportantInformation115
ToperforminitialsetupoftheMR400
Step Action
1PerformavisualinspectionoftheMR400,checkingforlooseor
missinghardwareordamage.
Iflooseormissinghardwareordamageisobserved,contact
technicalsupport.
2Installthemainbatteriesintothecartand
connectthereserve
batteries;seepage116.
3ConnectACmainspowertotheMR400,butDONOTturnonthe
powerswitch.Allowthebatteriestochargeforatleast12hours
beforeuse;seepage120.
Note
Before initial use, charge the batteries in the cart for at least 12
hours with the MR400 turned off and connected to AC mains power.
4IfwECGandwSpO2moduleswereincluded,performavisual
inspectionofthedevicesforlooseormissinghardwareordamage.
5IfawSpO2modulewasincluded,attachanSPO2probetoit;see
page118.
6Ifmodulebatterieswereincluded,chargethemodulebatteries
usingthePhilipsspecifiedbattery
charger.(Refertothe
instructionsprovidedwiththecharger.)
Note
Before initial use, charge the module batteries for at least 4 hours.
7Pressthepowerswitch(onthefrontofthecart)thenverifythatthe
MR400hassuccessfullypoweredupandthatthepowerLEDis
steadygreen.
Forotherpossibleindications,seepage27.
8IfequippedwithanIP5,verifythattheMR400andIP5aresettothe
same
wirelessnetworkchannelbycheckingthenetworkicon(see
examplesbelow)thatisdisplayed byeachdevice.
Seepage127forMR400networksettinginstructions;and,referto
theIFUfortheIP5fornetworksettinginstructions.
9IfwECGandwSpO2moduleswereincluded,installthecharged
modulebattery
(orbatteries)intothedevice(s);seepage124for
thewECGmoduleandseepage126forwSpO2module.