ExpressionMR400InstructionsforUse SpecificationsA7
HeartRate Automaticmodeselectsthevitalsigntoprovidetheheartrate
accordingtovitalsignsourceavailabilityandpriority.Ifnosource
available(ifnovitalsignmeetsthecriteria,thentheheartrate
sourcewillbedisplayedasNoneandno
heartratewillbe
produced.
Manualmodeselectiontoprovidetheheartrateisalsoavailable.
PulseOximeter Pulserate,pulsewaveform(normalized)andpercentsaturation
Trends Heartrate,respirationrate,P1and/orP2(systolic,diastolic,
mean),NIBP(systolic,diastolic,mean),EtCO2,O2,N2O,SpO2,
andAgents
CO2 BothendtidalandfractionalinspiredCO2
NIBP Pressures
(systolic,mean,diastolic)andstatus
RespirationRate Resp irationratederivedfrombellowsorCO2
N2O
EtN
2
OavailableinAgentMACbox
FiN
2
Onotdisplayed
O2 Inspired,endtidal(averagedpercent)
AGENT Automaticidentificationofprimaryandsecondaryagents
(Desflurane,Isoflurane,Enflurane,HalothaneorSevoflurane)
displayingbothendtidal(Et)andfractionalinspired(Fi)
concentrations.
Temperature Bodytemperature(°Cor°F)
ECG
ECG Amplifier
Protectedagainstdefibrillatorandelectrosurgerypotentials
StandardLeadConfigurations I,II,III,AVR,AVL,AVF
LeadFail Passive,sensingsignalimbalance
ECGInputImpedance >2.5M,singleended(accordingtoIEC60601227,50.102.3)
ElectrodeContactImpedance 20Kohms@10Hz
Heart Rate
Resolution 1beatperminute(BPM)
PulseRateRange 30to250BPM(Adult)
30to300BPM(Neonate,Pediatric)
Accuracy ±1percentor±1 BPM,whicheverisgreater.
Displayed Information