User Manual

Table Of Contents
ExpressionMR400InstructionsforUse SpecificationsA15
Transducer Adapter Cable Compatibility
InvasivepressureinputmateswithanAmphenolconnector(MS3106A14S6P).Withthisconnectorandthe
following connectioninformation,transduceradaptercabl esmaybefabricatedororderedfromvarious
manufacturers.
ConnectorPinNumber SignalName
A ‐Signal
B+Excitation
C+Signal
D ‐Excitation
EShield
AGENT (Optional)
SideStream,nondispersiveinfrared(NDIR)absorptiontechnique,includingwatertrapfiltrationsystemand
microprocessorlogiccontrolofsamplehandlingandcalibration
Simultaneouslymeasuredgases Anytwoofthefollowing,inspiredorexpired,whilealsomeasuring
CO2,N2O,and O2:
Halothane
Isoflurane
Sevoflurane
Desflurane
Enflurane
MeasurementRange(aftermaximumwarm
upperiod)
Halothane:0to5.0
Vol.%
Isoflurane:0to5.0Vol.%
Sevoflurane:0to8.0Vol.%
Desflurane:0to18.0Vol.%
Enflurane:0to5.0Vol.%
CarbonDioxide:0to10.0Vol.%
NitrousOxide:0to100Vol.%
Invasive Blood Pressure (Optional)